AMBU Automation Breathing Device
The U-Vent, AMBU or any standard breathing balloon is designed to automatically control you and perform invasive and non-invasive breathing.
U-Vent has a very sensitive mechanical system working with DC motor. Thanks to the sensitive position control sensors, the breathing is realized by squeezing and releasing the AMBU at the determined rates. There are 1 pressure sensor and two flow sensors in the device to control the gas flow, volume and pressure. Thanks to these sensitive sensors, the breathing process is precisely controlled.
U-Vent can optionally be used with an air-oxygen mixer. The desired oxygen rate can be adjusted in 10% slices.
U-Vent has a sensitive and large touch control screen in the size of 10". This
Device settings, graphics and real-time respiratory values can be read from the screen.Respirator, ventilator device, ambu automation, ventilation, respirator
Technicial Specifications
U-Vent is designed to be hung on the side barriers of patient beds, thanks to the hanger apparatus that can be attached and removed from the back of the device. Suspension apparatus can be simply attached and detached by hand with no tools
Safe breathing automation
.
The U-Vent AMBU has a protective cover made of transparent plexiglass for safety.
Thanks to the transparent structure of the cover, the operation of the system can be easily observed from the outside.
This cover prevents foreign matter from entering the work area, also interferes with AMBU
in order to prevent
they were designed.
In an emergency, the cover can be opened easily, allowing manual intervention.Technical document
Control :
Screen:
Ventilation modes:
Real-time values:
Ventilation parameters:
Network connection:
FiO2:
Graphics:
Alarms:
PLC
Oxygen - Air mixer (21-100%) (optional)
Respiratory rate 10-30 breaths/minute
Inspiration / Expiration ratio 1:1-
Inspiratory pause %0-50%
VCV (Volume Control Ventilation)
plateau pressure
Tidal Volume 100ml - 1000ml
10" touchscreen
PCV (Pressure Control Ventilation)
A/CV (Assist/Control Ventilation) (optional) CPAP (Non-Invasive) (optional)
1:3
Minimum airway pressure
Pause time (0-50% Ti)
Tidal volume
Maximum inspiratory pressure Plateau pressure
Minimum airway pressure
Volume Pressure Flow
Pressure limitation 10-60 cmH2O PEEP pressure 3-20 cmH2O High tidal volume
Low tidal volume Low airway pressure High airway pressure Apnea
Ethernet TCP/IP